Sample Record for 2021-03-07-2

Note: Fields with no data or with a value of "Unknown" will not be displayed on the print out.

Project Information

Project ID: 44444 DEA: ASCP EFIS: ASCP QA Service Level:
DIST-CO-RTE-PM: 59---
Work Description: Special project for submitting Sample information for Asphalt Supplier Certification Program.
Location Description:
Contractor Name: Unknown
Contractor Address: Unknown
Resident Engineer: Unknown Structure Rep.: Unknown

Sample Information

DIME Sample ID: 2021-03-07-2
TL Number from the TL Form:
Sample Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Sample Organization Contact: (916) 227-7302
Type of Sample: Verification
Type of Material: Asphalt Binder
Sample Taken Date: 03/07/2021
Date that the sample was received by the laboratory/organization performing the testing: 04/14/2021
The quantity of what the sample constitutes: 1 of 3 cans
A brief description of the sample: Asphalt Binder
From where was the sample collected: Manufacturer
Laboratory sample identification: 030721-826
Structure unique identifier:
Structure name:
The fabricator/manufacturer/facility name: VALERO - WILMINGTON

Material Information

Performance Grade: PG 70-10
Modified: No
Lot number:
Sublot number:
Tank number: 3A
Time of sample: 0800
HMA Supplier:
Certified Asphalt Supplier: Valero Marketing & Supply - Wilmington CA
Multigrade:
Multigrade Type:
Asphalt Binder Mix Number:

Test Results

Test 1 of 6

DIME Test ID: 2021-03-07-2-1
Test Method: AASHTO T240-17: Standard Method of Test for Effect of Heat and Air on a Moving Film of Asphalt Binder (Rolling Thin-Film Oven Test)
Verified Date: 04/21/2021
Date Tested: 04/15/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: as per 2018 standard specs
Test Result Compliance: complies
Additional Comments When Verified:
Is mass change based on one or two containers? Two
Any material loss from containers during heating process/testing? No
Mass of empty container #1 181.94 grams
Mass of empty container #2 166.2 grams
Mass of container #1 before test (m1i) 217.817 grams
Mass of container #2 before test (m2i) 201.789 grams
Mass of container #1 after test (m1f) 217.762 grams
Mass of container #2 after test (m2f) 201.73 grams
Average mass change -0.16 percent

Test 2 of 6

DIME Test ID: 2021-03-07-2-2
Test Method: AASHTO T51-13: Standard Method of Test for Ductility of Asphalt Materials (Asphalt Binder)
Verified Date: 04/21/2021
Date Tested: 04/15/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: as per 2018 standard specs
Test Result Compliance: complies
Additional Comments When Verified:
Is a normal test obtainable? Yes
Number of replicates tested Two
Ductility of test #1 (D1) 90 cm
Ductility of test #2 (D2) 97 cm
Ductility of sample 94 cm

Test 3 of 6

DIME Test ID: 2021-03-07-2-3
Test Method: AASHTO T315-16: Standard Method of Test for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer (DSR) (Dynamic Shear)
Verified Date: 04/21/2021
Date Tested: 04/15/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: per 2018 standard specifications ORIG
Test Result Compliance: complies
Additional Comments When Verified:
Phase Original binder
Test Plate Diameter 25 mm
Test Gap 1 mm
Test Frequency 10.0 rad/s
Strain Amplitude 12.00 %
Stress Amplitude 0.146 kPa
Phase Angle (δ) (for the 10 cycles) 86.9 degree
Complex Modulus (G*) (for the 10 measurements) 1.214 kPa
Test Temperature (Specimen 1) 70.0 C
Dynamic Shear, G*/sin(δ) 1.22 kPa

Test 4 of 6

DIME Test ID: 2021-03-07-2-4
Test Method: AASHTO T315-16: Standard Method of Test for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer (DSR) (Dynamic Shear)
Verified Date: 04/21/2021
Date Tested: 04/15/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: Per 2018 standard specifications DSR RTFO
Test Result Compliance: complies
Additional Comments When Verified:
Phase RTFO residue
Test Plate Diameter 25 mm
Test Gap 1 mm
Test Frequency 10.0 rad/s
Strain Amplitude 10.00 %
Stress Amplitude 0.312 kPa
Phase Angle (δ) (for the 10 cycles) 83.3 degree
Complex Modulus (G*) (for the 10 measurements) 3.117 kPa
Test Temperature (Specimen 1) 70.0 C
Dynamic Shear, G*/sin(δ) 3.14 kPa

Test 5 of 6

DIME Test ID: 2021-03-07-2-5
Test Method: AASHTO T315-16: Standard Method of Test for Determining the Rheological Properties of Asphalt Binder Using a Dynamic Shear Rheometer (DSR) (Dynamic Shear)
Verified Date: 04/22/2021
Date Tested: 04/21/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: Per 2018 Std Specs PAV
Test Result Compliance: complies
Additional Comments When Verified:
Phase PAV residue
Test Plate Diameter 8 mm
Test Gap 2 mm
Test Frequency 10.0 rad/s
Strain Amplitude 1.00 %
Stress Amplitude 22.254 kPa
Phase Angle (δ) (for the 10 cycles) 51.1 degree
Complex Modulus (G*) (for the 10 measurements) 2225.363 kPa
Test Temperature (Specimen 1) 34.0 C
(G*) x sin(δ)= 1732 kPa

Test 6 of 6

DIME Test ID: 2021-03-07-2-6
Test Method: AASHTO T313-19: Standard Method of Test for Determining the Flexural Creep Stiffness of Asphalt Binder Using the Bending Beam Rheometer (BBR)
Verified Date: 04/22/2021
Date Tested: 04/20/2021
Submitted By Organization: CT HQ - Sacramento Asphalt Binder Laboratory
Organization Address: 5900 Folsom Boulevard, Room 225,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: guadalupe.magana@dot.ca.gov
Organization Phone: (916) 227-7302
General Comments for this Test: BBR
Test Result Compliance: complies
Additional Comments When Verified:
Time test started no data HH:MM:SS
Flags issued by software during test no data
Correlation coefficient (R²) no data0
Regression coefficient a no data
Regression coefficient b no data
Regression coefficient c no data
Load at time 0.0 no data0 mN
Deflection at time 0.0 s no data0 µm
Load at time 0.5 s no data0 mN
Deflection at time 0.5 s no data0 µm
Load at 60.0 s 989 mN
Beam deflection 60.0 s 1.208 µm
Measured Stiffness Modulus 60.0 s 66 mPa
Estimated Stiffness Modulus 60.0 s 66 mPa
Difference between stiffness moduli 60.0 s no data0 %
Estimated m-value 60.0 s 0.411