Project ID: 8801 DEA: AML-CEM EFIS: AML-CEM
QA Service Level:
DIST-CO-RTE-PM: 59---
Work Description: Special project for Cementitious Materials for use in Concrete Authorized Material List (AML)
Location Description:
Contractor Name: Unknown
Contractor Address: Unknown
Resident Engineer: Unknown Structure Rep.: Unknown
DIME Sample ID: 2020-09-01-20
TL Number from the TL Form:
Sample Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Sample Organization Contact: (916) 227-7071
Type of Sample: Proficiency
Type of Material: Cementitious
Sample Taken Date: 09/01/2020
Date that the sample was received by the laboratory/organization performing the testing: 09/01/2020
The quantity of what the sample constitutes: 1 bag
A brief description of the sample: pozzolan
From where was the sample collected: Other
Laboratory sample identification: CM208067
Structure unique identifier:
Structure name:
The fabricator/manufacturer/facility name: CCRL
DIME Test ID: 2020-09-01-20-2
Test Method: ASTM C109-16a: Standard Test Method for Compressive Strength of Hydraulic Cement Mortars (Using 2-in. or [50-mm] Cube Specimens)
Verified Date: 07/11/2022
Date Tested: 09/22/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test:
Test Result Compliance:
Additional Comments When Verified:
Will control cubes be used per ASTM C311?
|
yes
|
Age of specimen
|
7 days
|
The amount of flow
|
132 %
|
The amount of water used
|
47 %
|
Compressive strength of 1st specimen
|
4830 psi
|
Compressive strength of 2nd specimen
|
4851 psi
|
Compressive strength of 3rd specimen
|
4802 psi
|
Average Compressive strength (to nearest 10 psi)
|
4830 psi
|
Average Compressive strength (to nearest 10 psi) control group
|
4850 psi
|
Strength Activity Index of Portland Cement per ASTM C109
|
100 %
|
DIME Test ID: 2020-09-01-20-3
Test Method: ASTM C109-16a: Standard Test Method for Compressive Strength of Hydraulic Cement Mortars (Using 2-in. or [50-mm] Cube Specimens)
Verified Date: 07/11/2022
Date Tested: 10/13/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test:
Test Result Compliance:
Additional Comments When Verified:
Will control cubes be used per ASTM C311?
|
yes
|
Age of specimen
|
28 days
|
The amount of flow
|
132 %
|
The amount of water used
|
47 %
|
Compressive strength of 1st specimen
|
6113 psi
|
Compressive strength of 2nd specimen
|
6473 psi
|
Compressive strength of 3rd specimen
|
6488 psi
|
Average Compressive strength (to nearest 10 psi)
|
6360 psi
|
Average Compressive strength (to nearest 10 psi) control group
|
5700 psi
|
Strength Activity Index of Portland Cement per ASTM C109
|
112 %
|
DIME Test ID: 2020-09-01-20-4
Test Method: ASTM C1038-14b: Standard Test Method for Expansion of Hydraulic Cement Mortar Bars Stored in Water (Cement Mortar Bars)
Verified Date: 07/11/2022
Date Tested: 09/16/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test: Only 3 bars fabricated. Specimen D numbers are chosen to make the average correct.
Test Result Compliance:
Additional Comments When Verified:
Specimen A Initial measurement at 24 hours
|
0.268 in
|
Specimen B Initial measurement at 24 hours
|
0.2807 in
|
Specimen C Initial measurement at 24 hours
|
0.2738 in
|
Specimen D Initial measurement at 24 hours
|
0.0087 in
|
Specimen A Final measurement at 14 days
|
0.2592 in
|
Specimen B Final measurement at 14 days
|
0.272 in
|
Specimen C Final measurement at 14 days
|
0.2651 in
|
Specimen D Final measurement at 14 days
|
no data0 in
|
The average length of change of the four specimens
|
-0.087 %
|
DIME Test ID: 2020-09-01-20-5
Test Method: ASTM C430-8: Standard Test Method for Fineness of Hydraulic Cement by the 45-μm (No. 325) Sieve
Verified Date: 07/11/2022
Date Tested: 10/20/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test:
Test Result Compliance:
Additional Comments When Verified:
Correction Factor (CF)
|
.39
|
Corrected Residue (Rc)
|
12.9
|
DIME Test ID: 2020-09-01-20-6
Test Method: ASTM C151-15: Standard Test Method for Autoclave Expansion of Hydraulic Cement
Verified Date: 07/11/2022
Date Tested: 10/20/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test:
Test Result Compliance:
Additional Comments When Verified:
Initial length of specimen
|
0.2929
|
Final length of specimen
|
0.2942
|
Change in length of specimen(In percent)
|
0.01 %
|
Gauge Length
|
10
|
DIME Test ID: 2020-09-01-20-8
Test Method: ASTM C188-15: Standard Test Method for Density of Hydraulic Cement
Verified Date: 07/11/2022
Date Tested: 10/20/2020
Submitted By Organization: CT HQ - Sacramento Cement Testing Laboratory
Organization Address: 5900 Folsom Boulevard,
Organization City: Sacramento
Organization State: CA
Organization Zipcode: 95819
Organization Email: ben.lenz@dot.ca.gov
Organization Phone: (916) 227-7071
General Comments for this Test:
Test Result Compliance:
Additional Comments When Verified:
Density of hydraulic cement
|
2.65 g/cm^3
|